HomeCategoria di prodottoEMI filterMRI Filter F1770X

EMI filter

MRI Filter F1770X Serie

Filter

MRI Room single line filters deliver clean power to MRI scanning equipment to prevent interference with their performance. The filters are designed with minimum leakage current to ground and provide 100dB attenuation minimum from 5MHz to 10GHz and the current up to 50Amp.

    • MRI Room single line filters deliver clean power to MRI scanning equipment to prevent interference with their performance. The filters are designed with minimum leakage current to ground and provide 100dB attenuation minimum from 5MHz to 10GHz.
    • Voltage Rating: 120VAC/277VAC 
    • Frequency: 50/60Hz
    • Hi-Pot: Line to Ground: 2250VDC for 2 sec       
                  Line to Line: 1450VDC for 2 sec
    • Ambient Temperature Range: -25°C To +85°C
    • Filter components are encapsulated in a plated metal enclosure
    • Leakage Current: 50mA max. @ 250VAC 60Hz
    • UL and cUL listed

Caratteristiche elettriche della MRI Filter F1770X Serie.

Codice

Descrizione della parte

top